Acquisition of ground calibration data from the AXAF CCD Imaging Spectrometer, one of two focal plane instruments on NASA's Advanced X-ray Astrophysics Facility, was completed in 1997. Here we summarize results of the detector level calibration effort. Our calibration program has included measurements of CCD response to undispersed synchrotron radiation, measurements of x-ray absorption fine structure, and of sub-pixel structure in the detector. Errors in the energy scale are at the level of a few tenths of one percent, and detection efficiency errors are no large than a few percent. We have also obtained new insights into the mechanisms by which the CCD gate structure and channel stops influence the CCD spectral redistribution function.
Baganoff Fred
Bautz Mark W.
Grant Catherine E.
Isobe Takashi
Jones Stephen E.
No associations
LandOfFree
X-ray CCD calibration for the AXAF CCD Imaging Spectrometer does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with X-ray CCD calibration for the AXAF CCD Imaging Spectrometer, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and X-ray CCD calibration for the AXAF CCD Imaging Spectrometer will most certainly appreciate the feedback.