The operational problems of the gamma ray spectrometer shielded passively with 12 cm of lead and actively by five 0.5 m × 0.5 m × 0.05 m plastic veto shields are described. In the present paper we describe the new results of our ground based actively shielded extended range low level gamma spectrometer emphasizing the anticoincidence timing details which are unavoidable for the proper operation of the system After the new adjustment of active shield germanium system, total count rate in anticoincidence background spectrum was 0.38 c/s (50 keV-2 MeV), the achieved reduction factor with respect to the passively shield background spectrum is 2.42. The 511 keV reduction factor is 4.25.
Bikit Ištvan
Forkapić Sofija
Mrđa Dušan
Slivka Jaroslav
Todorovic Nataša
No associations
LandOfFree
Fast Timed Active Shield for a Gamma-Ray Spectrometer does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Fast Timed Active Shield for a Gamma-Ray Spectrometer, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Fast Timed Active Shield for a Gamma-Ray Spectrometer will most certainly appreciate the feedback.