The design and performance of an electron spin resonance spectrometer operating at 3 and 9 GHz microwave frequencies combined with a 9 T superconducting magnet is described. The probehead contains a compact two-loop, one gap resonator and is embedded in the variable temperature insert of the magnet enabling measurements in the 0- 9 T magnetic field and 1.5-400 K\ temperature range. The spectrometer allows studies on systems where resonance occurs at fields far above the $g$ $\approx 2$ paramagnetic condition such as in strongly interacting spin systems. The low quality factor of the resonator allows time resolved experiments such as e.g. longitudinally detected ESR. We demonstrate the performance of the spectrometer on the MgB$_{2}$ superconductor and the RbC$_{60}$ conducting alkaline fulleride polymer.
Muranyi Ferenc
Simon Ferenc
No associations
LandOfFree
ESR spectrometer with a loop-gap resonator for cw and time resolved studies in a superconducting magnet does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with ESR spectrometer with a loop-gap resonator for cw and time resolved studies in a superconducting magnet, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and ESR spectrometer with a loop-gap resonator for cw and time resolved studies in a superconducting magnet will most certainly appreciate the feedback.