Energy resolution ΔE=12 eV at 5.9 keV for lead absorber coupled to a single Al-STJ via phonons

Physics

Scientific paper

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

1

Superconducting Optical, X-Ray, And Gamma-Ray Detectors, X- And Gamma-Ray Sources, Mirrors, Gratings, And Detectors

Scientific paper

An energy resolution of ΔE=12 eV (FWHM) has been measured for the 55Mn Kα1-line (E=5.9 keV) using a superconducting lead absorber (90×90×1.3 μm3) read out by a single aluminum superconducting tunnel junction (100×100 μm2). The total electronic noise contribution was measured as ΔEelec=4 eV (FWHM). The superconducting tunnel junction is located on a Si3N4-membrane of thickness 0.3 μm, the lead absorber is separated from the superconducting tunnel junction's top layer by a thin layer of natural aluminum oxide. .

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

Energy resolution ΔE=12 eV at 5.9 keV for lead absorber coupled to a single Al-STJ via phonons does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with Energy resolution ΔE=12 eV at 5.9 keV for lead absorber coupled to a single Al-STJ via phonons, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Energy resolution ΔE=12 eV at 5.9 keV for lead absorber coupled to a single Al-STJ via phonons will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFWR-SCP-O-1790003

  Search