We present the first operation of the Avalanche Photodiode (APD) from Hamamatsu to xenon scintillation light and to direct X-rays of 22.1 keV and 5.9 keV. A large non-linear response was observed for the direct X-ray detection. At 415 V APD bias voltage it was of about 30 % for 22.1 keV and about 45 % for 5.9 keV. The quantum efficiency for 172 nm photons has been measured to be 69 +/- 15 %.
Amaro F. D.
Ballester Otger
Freitas E. D. C.
Jover-Manas G. V.
Lux Thomas
No associations
LandOfFree
Characterization of the Hamamatsu S8664 Avalanche Photodiode for X-Ray and VUV-light detection does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Characterization of the Hamamatsu S8664 Avalanche Photodiode for X-Ray and VUV-light detection, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Characterization of the Hamamatsu S8664 Avalanche Photodiode for X-Ray and VUV-light detection will most certainly appreciate the feedback.