Characterization of the Hamamatsu S8664 Avalanche Photodiode for X-Ray and VUV-light detection

Physics – Instrumentation and Detectors

Scientific paper

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

11 pages, 3 figures, submitted to Elsevier

Scientific paper

We present the first operation of the Avalanche Photodiode (APD) from Hamamatsu to xenon scintillation light and to direct X-rays of 22.1 keV and 5.9 keV. A large non-linear response was observed for the direct X-ray detection. At 415 V APD bias voltage it was of about 30 % for 22.1 keV and about 45 % for 5.9 keV. The quantum efficiency for 172 nm photons has been measured to be 69 +/- 15 %.

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

Characterization of the Hamamatsu S8664 Avalanche Photodiode for X-Ray and VUV-light detection does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with Characterization of the Hamamatsu S8664 Avalanche Photodiode for X-Ray and VUV-light detection, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Characterization of the Hamamatsu S8664 Avalanche Photodiode for X-Ray and VUV-light detection will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFWR-SCP-O-319381

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.