Physics – Instrumentation and Detectors
Scientific paper
2009-05-15
Astropart.Phys.33:13-18,2010
Physics
Instrumentation and Detectors
15 pages, 6 figures
Scientific paper
10.1016/j.astropartphys.2009.10.
A method is described that allows calibration and assessment of the linearity of response of an array of photomultiplier tubes. The method does not require knowledge of the photomultiplier single photoelectron response model and uses science data directly, thus eliminating the need for dedicated data sets. In this manner all photomultiplier working conditions (e.g. temperature, external fields, etc) are exactly matched between calibration and science acquisitions. This is of particular importance in low background experiments such as ZEPLIN-III, where methods involving the use of external light sources for calibration are severely constrained.
Akimov Yu. D.
Araújo Henrique M.
Barnes Joshua E.
Belov Anatoly V.
Burenkov Alexander
No associations
LandOfFree
Calibration of Photomultiplier Arrays does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Calibration of Photomultiplier Arrays, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Calibration of Photomultiplier Arrays will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-524362