Additional correction for energy transfer efficiency calculation in filter-based Förster resonance energy transfer microscopy for more accurate results

Physics – Optics

Scientific paper

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

1

Band-Pass Filters, Biological Techniques, Bio-Optics, Fluorescence, Molecular Biophysics, Optical Filters, Optical Microscopy, Proteins, Biological Techniques And Instrumentation, Biomedical Engineering, Optical Microscopy, Proteins, Conventional Optical Microscopes, Filters, Zone Plates, And Polarizers

Scientific paper

Förster resonance energy transfer (FRET) microscopy is commonly used to monitor protein interactions with filter-based imaging systems, which require spectral bleedthrough (or cross talk) correction to accurately measure energy transfer efficiency (E). The double-label (donor+acceptor) specimen is excited with the donor wavelength, the acceptor emission provided the uncorrected FRET signal and the donor emission (the donor channel) represents the quenched donor (qD), the basis for the E calculation. Our results indicate this is not the most accurate determination of the quenched donor signal as it fails to consider the donor spectral bleedthrough (DSBT) signals in the qD for the E calculation, which our new model addresses, leading to a more accurate E result. This refinement improves E comparisons made with lifetime and spectral FRET imaging microscopy as shown here using several genetic (FRET standard) constructs, where cerulean and venus fluorescent proteins are tethered by different amino acid linkers.

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

Additional correction for energy transfer efficiency calculation in filter-based Förster resonance energy transfer microscopy for more accurate results does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with Additional correction for energy transfer efficiency calculation in filter-based Förster resonance energy transfer microscopy for more accurate results, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Additional correction for energy transfer efficiency calculation in filter-based Förster resonance energy transfer microscopy for more accurate results will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFWR-SCP-O-883831

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.