Physics
Scientific paper
Feb 1996
adsabs.harvard.edu/cgi-bin/nph-data_query?bibcode=1996nimpa.376..174g&link_type=abstract
Nuclear Instruments and Methods in Physics Research Section A, v. 376, p. 174-184.
Physics
16
Scientific paper
A recoil separator in combination with a windowless gas target has been designed for the measurement of the radiative capture reaction p(7Be, γ)8B. The separator consists of momentum and velocity filters and a ΔE-E detector telescope. The setup was tested quantitatively using the p(12C, γ)13N reaction at the effective energy Ecm = 841 keV. Projectile fluxes were measured directly with Faraday cups and indirectly with elastic backscattering into Si detectors, while the 13N recoil flux was measured directly with the ΔE-E telescope. A suppression of the 12C beam particles by a factor 2 × 10-10 was observed when the system was tuned for the recoil 13N's. Special emphasis was given to the charge state probabilities of the 13N recoils. Possible improvements of the system for the measurement of other capture reactions are discussed.
Brand Kate
Campajola L.
D'Onofrio Alberto
Gialanella Lucio
Greife Uwe
No associations
LandOfFree
A recoil separator for the measurement of radiative capture reactions does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with A recoil separator for the measurement of radiative capture reactions, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and A recoil separator for the measurement of radiative capture reactions will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-855254