Physics – Condensed Matter – Other Condensed Matter
Scientific paper
2005-03-04
Physics
Condensed Matter
Other Condensed Matter
Scientific paper
10.1016/j.jmmm.2005.04.014
The long-time temperature drift in a commercial Quantum Design MPMS SQUID magnetometer was evaluated using time-dependent magnetization measurements of Gd$_2$O$_3$. In contrast to earlier claims, the amplitude of the drift was found not to exceed 1-1.5 K. 30 minutes after system stabilization the temperature deviation did not exceed 0.2 K and the temperature was fully stabilized in less than 3 hours.
Bud'ko Sergey L.
Canfield Paul. C.
No associations
LandOfFree
Evaluation of a long-time temperature drift in a commercial Quantum Design MPMS SQUID magnetometer using Gd$_2$O$_3$ as a standard does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Evaluation of a long-time temperature drift in a commercial Quantum Design MPMS SQUID magnetometer using Gd$_2$O$_3$ as a standard, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Evaluation of a long-time temperature drift in a commercial Quantum Design MPMS SQUID magnetometer using Gd$_2$O$_3$ as a standard will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-200872