adsabs.harvard.edu/cgi-bin/nph-data_query?bibcode=1990spie.1341..268m&link_type=abstract
IN: Infrared technology XVI; Proceedings of the Meeting, San Diego, CA, July 11-13, 1990 (A92-21426 07-35). Bellingham, WA, Soci
Physics
Absorption Spectra, Far Infrared Radiation, Infrared Absorption, Methane, Molecular Spectra, Ray Tracing, Reflectance
Scientific paper
A mirrorless multiple path cell which can be efficiently used in the FIR is described. The device is based on the White cell and light pipe methods, and requires no setup even after cooling. It is shown how the system can be efficiently operated if certain geometrical conditions are maintained which ensure a uniform light energy distribution inside the cell. The optimization of some physical parameters is discussed, and it is shown that an effective absorption pathlength as long as 4 or 5 m can be obtained with a simple absorption cell which is 10 cm long. The FIR spectrum of CH4 obtained using the cell is presented.
Marteau Ph.
Obriot J.
No associations
LandOfFree
Far infrared multiple-path cell without internal mirrors does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Far infrared multiple-path cell without internal mirrors, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Far infrared multiple-path cell without internal mirrors will most certainly appreciate the feedback.