Laser metrology gauges for OSI

Physics

Scientific paper

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

9

Scientific paper

Heterodyne interferometers have been commercially available for many years. In addition, many versions have been built at JPL for various projects. This activity is aimed at improving the accuracy of such interferometers from the 1 - 30 nanometer level to the picometer level for use in the proposed OSI and SONATA missions as metrology gauges. In the null-gauge configuration, we obtained a precision of 0.6 picometers at time scales of 2,500 seconds. In the relative-gauge configuration, we obtained an accuracy of 3.5 picometers rms in vacuum at time scales of few minutes. As absolute gauge with an accuracy of 10 microns over a distance of 10 meters in under construction.

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

Laser metrology gauges for OSI does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with Laser metrology gauges for OSI, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Laser metrology gauges for OSI will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFWR-SCP-O-1543201

  Search