The Four-Quadrant Phase Mask Coronagraph. III. Laboratory Performance

Physics – Optics

Scientific paper

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

28

Instrumentation: Adaptive Optics, Methods: Laboratory, Planetary Systems, Techniques: Image Processing

Scientific paper

We report a laboratory experiment to assess the performance and limitations of the four-quadrant phase mask coronagraph we proposed 2 years ago. A total flux rejection factor of 4400 was reached using off-the-shelf optical components. The peak intensity of the Airy pattern is correspondingly reduced by a factor of 44,000 with a stability of a few hours. In this paper, we discuss implications of this excellent result regarding the next generation of instruments for extrasolar planet detection.

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

The Four-Quadrant Phase Mask Coronagraph. III. Laboratory Performance does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with The Four-Quadrant Phase Mask Coronagraph. III. Laboratory Performance, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and The Four-Quadrant Phase Mask Coronagraph. III. Laboratory Performance will most certainly appreciate the feedback.