Physics
Scientific paper
Aug 2008
adsabs.harvard.edu/cgi-bin/nph-data_query?bibcode=2008wfc..rept...25h&link_type=abstract
Instrument Science Report WFC3 2008-25, 16 pages
Physics
Hubble Space Telescope, Hst, Space Telescope Science Institute, Wfc3, Wide Field Camera 3, Thermal Vacuum 3, Ir Channel, Read Noise
Scientific paper
Using data taken during WFC3's Thermal Vacuum 3 (TV3) testing campaign, we have characterized the readnoise behavior in the IR Channel, which contained IR-4 (FPA165). The Contract End Item (CEI) Specifications call for a correlated double sampling (CDS) readnoise of 15 e-, along with an achievable readnoise of 10 e- from a ramp of 10 nondestructive reads. We observe a CDS readnoise of roughly 20.3 - 21.0 e-, and an effective readnoise of 8.5 - 17 e- for 16 non-destructive reads, depending on sample sequence. The IR channel meets the spec for achievable readnoise only when using the MIF sample sequences and a line-fitting data reduction technique.
No associations
LandOfFree
WFC3 TV3 Testing: IR Channel Read Noise does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with WFC3 TV3 Testing: IR Channel Read Noise, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and WFC3 TV3 Testing: IR Channel Read Noise will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-1064512