Physics
Scientific paper
Nov 2009
adsabs.harvard.edu/cgi-bin/nph-data_query?bibcode=2009wfc..rept...20s&link_type=abstract
Instrument Science Report WFC3 2009-20, 9 pages
Physics
Hubble Space Telescope, Hst, Space Telescope Science Institute, Wide Field Camera 3, Wfc3
Scientific paper
The image quality analysis carried out during the third thermal-vacuum campaign (TV3) revealed that vibrations of the UVIS shutter cause changes in the width and in the central pixel flux of point sources in short exposures. During the Servicing Mission Observatory Verification (May-August 2009) we observed the flux standard star GD153 with a series of 0.5 sec, 1.0 sec, 3.0 sec, 10.0 sec and 20.0 sec exposures to investigate the impact of the shutter vibration on data acquired in orbit. Our analysis shows that in the very short exposures (<3.0-10.0 sec) the PSF is significantly degraded, but that the flux is preserved.
No associations
LandOfFree
WFC3 SMOV Program 11798: UVIS PSF Core Modulation does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with WFC3 SMOV Program 11798: UVIS PSF Core Modulation, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and WFC3 SMOV Program 11798: UVIS PSF Core Modulation will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-1152631