Physics – Instrumentation and Detectors
Scientific paper
2011-05-04
Rev. Sci. Instrum. 82, 093110 (2011)
Physics
Instrumentation and Detectors
Fixed a broken figure, updated references. 6 pages, 7 figures, 1 table
Scientific paper
10.1063/1.3641294
We report operation and characterization of a lab-assembled single-photon detector based on commercial silicon avalanche photodiodes (PerkinElmer C30902SH, C30921SH). Dark count rate as low as 5 Hz was achieved by cooling the photodiodes down to -80 C. While afterpulsing increased as the photodiode temperature was decreased, total afterpulse probability did not become significant due to detector's relatively long deadtime in a passively-quenched scheme. We measured photon detection efficiency higher than 50% at 806 nm.
Jeong Youn-Chang
Kim Yong-Su
Kim Yoon-Ho
Makarov Valeri V.
Sauge Sebastien
No associations
LandOfFree
Ultra-low noise single-photon detector based on Si avalanche photodiode does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Ultra-low noise single-photon detector based on Si avalanche photodiode, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Ultra-low noise single-photon detector based on Si avalanche photodiode will most certainly appreciate the feedback.