Physics
Scientific paper
Nov 2006
adsabs.harvard.edu/cgi-bin/nph-data_query?bibcode=2006nimpa.568...72l&link_type=abstract
Nuclear Instruments and Methods in Physics Research Section A, Volume 568, Issue 1, p. 72-77.
Physics
1
Tracking And Position-Sensitive Detectors, Solid-State Detectors, Elemental Semiconductors
Scientific paper
We report on test beam results obtained with a 32.5 cm2 microstrip detector processed on an n-type 380 μm thick magnetic Czochralski (MCZ) grown silicon substrate with 1200 Ωcm effective resistivity. The full depletion voltage of the as-processed detector was 420 V with a leakage current of 2 μA. The AC coupled detector had 1024 p+ strips, 10 μm by width and 6.154 cm by length with a pitch of 50 μm. The detector was irradiated with 10 MeV protons to 1.6×1014 1 MeV neutron equivalent fluence and annealed for 345 days at room temperature. The post-irradiation full depletion voltage of the detector was 225 V. The leakage current at the full depletion measured at -10 °C was 261 μA. The beam tests were carried out at the CERN H2 area using a Silicon Beam Telescope, which consists of pairs of horizontal and vertical position sensitive silicon detectors. This telescope determines the tracks of incoming particles and hence provides a reference measurement for the detector characterization. In the beam test an average signal to noise ratio of 3 with a spatial resolution of 20 μm and a particle detection efficiency of 36% were measured. These results show that the MCZ device detected particles, which encourages further investigations of MCZ silicon as a detector material. The poor performance of the MCZ detector may be explained by the problems observed in the reference telescope.
Czellar Sandor
Härkönen Jaakko
Hæggström Edward
Karimaki V.
Kassamakov Ivan
No associations
LandOfFree
Test beam results of a proton irradiated Czochralski silicon strip detector does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Test beam results of a proton irradiated Czochralski silicon strip detector, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Test beam results of a proton irradiated Czochralski silicon strip detector will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-1344539