Physics – Nuclear Physics – Nuclear Experiment
Scientific paper
2003-11-24
Nucl.Instrum.Meth. A527 (2004) 512-518
Physics
Nuclear Physics
Nuclear Experiment
submitted to Nuclear Instruments and Methods in Phys. Res. Sect. A
Scientific paper
10.1016/j.nima.2004.03.179
The study of the neutron response and n-gamma discrimination for small (18x26x8) mm3 liquid scintillator BC501A (Bicron) detector was carried out by digital charge comparison method. Three ranges of neutron energies were used: uniform distribution from 0.95 MeV to 1.23 MeV, continuous spectra of AmBe source and monoenergetic 16.2 MeV neutrons. The obtained results are compared with those for cylindrical liquid scintillation detector (40 mm diameter, 60 mm length) at the same energies of neutrons. A dramatic fall of the neutron response function at 400 keVee for small detector at 16.2 MeV neutron energy was measured. For (0.95 - 1.23) MeV neutron energy range such fall takes place at 260 keVee. The greater slope of neutron locus at (0.95 - 1.23) MeV neutron energy comparing to 16.2 MeV for both detectors is explained by longer tail of pulse from proton recoils within (0.1-1.23) MeV energy range.
Cerny Jiri
Dolezal Zdenek
Ivanov P. M.
Kuzmin E. S.
Svejda J.
No associations
LandOfFree
Study of neutron response and n-gamma discrimination by charge comparison method for small liquid scintillation detector does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Study of neutron response and n-gamma discrimination by charge comparison method for small liquid scintillation detector, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Study of neutron response and n-gamma discrimination by charge comparison method for small liquid scintillation detector will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-670968