Physics – Instrumentation and Detectors
Scientific paper
2010-02-14
Physics
Instrumentation and Detectors
14 pages, 5 figures
Scientific paper
10.1016/j.nima.2010.03.151
Oxygen, nitrogen and methane purification efficiencies for a common zirconium getter are measured in 1050 Torr of xenon gas. Starting with impurity concentrations near 10^{-6} g/g, the outlet impurity level is found to be less than 120*10^{-12} g/g for O2 and less than 950*10^{-12} g/g for N2. For methane we find residual contamination of the purified gas at concentrations varying over three orders of magnitude, depending on the purifier temperature and the gas flow rate. A slight reduction in the purifier's methane efficiency is observed after 13 mg of this impurity has been absorbed, which we attribute to partial exhaustion of the purifier's capacity for this species. We also find that the purifier's ability to absorb N2 and methane can be extinguished long before any decrease in O2 performance is observed, and slower flow rates should be employed for xenon purification due to the cooling effect that the heavy gas has on the getter.
Dobi A.
Hall Carter
Kaufman Larry
Langford T.
Leonard Douglas Sidney
No associations
LandOfFree
Study of a zirconium getter for purification of xenon gas does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Study of a zirconium getter for purification of xenon gas, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Study of a zirconium getter for purification of xenon gas will most certainly appreciate the feedback.