Studies of 100 um-thick silicon strip detector with analog VLSI readout

Physics – Instrumentation and Detectors

Scientific paper

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

8 pages, 6 figures, Submitted to Nucl. Instr. Meth. A

Scientific paper

We evaluate the performances of a 100 um-thick silicon strip detector (SSD) with a 300 MeV proton beam and a 90Sr beta-ray source. Signals from the SSD have been read out using a VLSI chip. Common-mode noise, signal separation efficiency and energy resolution are compared with those for the SSD's with a thickness of 300 um and 500 um. Energy resolution for minimum ionizing particles (MIP's) is improved by fitting the non-constant component in a common-mode noise with a linear function.

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

Studies of 100 um-thick silicon strip detector with analog VLSI readout does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with Studies of 100 um-thick silicon strip detector with analog VLSI readout, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Studies of 100 um-thick silicon strip detector with analog VLSI readout will most certainly appreciate the feedback.