Aug 2005
adsabs.harvard.edu/cgi-bin/nph-data_query?bibcode=2005spie.5922...56b&link_type=abstract
Hard X-Ray and Gamma-Ray Detector Physics VII. Edited by James, Ralph B.; Franks, Larry A.; Burger, Arnold. Proceedings of the
Physics
Scientific paper
The stability of high-pressure xenon gamma-ray specrometers has been measured at Mirmar Sensor. Stability was measured by tracking the 137Cs peak of the entire spectroscopy system including detector, preamplifier, amplifier, and multichannle analyzer over 6 days in a laboratory setting. Overall system gain drift is about 0.016%. Resolution drifts are about 0.15%. The overall energy resolution is between 1.7 and 1.8% at 662 keV.
No associations
LandOfFree
Stability of high-pressure xenon gamma-ray spectrometers does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Stability of high-pressure xenon gamma-ray spectrometers, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Stability of high-pressure xenon gamma-ray spectrometers will most certainly appreciate the feedback.