Physics
Scientific paper
Jan 2006
adsabs.harvard.edu/cgi-bin/nph-data_query?bibcode=2006e%26psl.241..381b&link_type=abstract
Earth and Planetary Science Letters, Volume 241, Issue 3-4, p. 381-386.
Physics
5
Scientific paper
Manual reorientation of traditional 10.5 cm3 paleomagnetic cores between successive spins in Molspin-type magnetometers produces some operator error that may be significantly reduced if smaller core (5.1 cm3) is used instead. The smaller core may be locked into a cubical holder that ensures its precise angular alignment between successive spins of the magnetometer. This significantly improves precision of the mean vector direction for replicate measurements, or for a suite of specimens, provided that the magnetic mineralogy of the rock is sufficiently homogenous to be meaningfully characterized in the smaller (5.1 cm3) core. The smaller core is more quickly measured and less expensive to drill.
Almqvist B. S.
Borradaile Graham J.
Lucas Kevin
No associations
LandOfFree
Specimen size and improved precision with the Molspin spinner magnetometer does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Specimen size and improved precision with the Molspin spinner magnetometer, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Specimen size and improved precision with the Molspin spinner magnetometer will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-1203547