Simulation of the Avalanche Process in the G-APD and Circuitry Analysis of the SiPM

Physics – Instrumentation and Detectors

Scientific paper

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

9 pages, 5 figures, pdf

Scientific paper

The discrete modeling of the Geiger-mode APD is considered. Results of modeling and experimental measurements with the SiPM show that the known formula for the charge of the avalanche pulse Q=dU*Cd underestimates its value. In addition, it is seen from the dynamic of the avalanche multiplication that the resistor Rq in photodiode, usually called a quenching resistor, in reality fulfills only the restoring function. The SiPM restoring time, taken into account the number of pixel N and the load resistance R, is T=Cd*(Rq+NR).

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

Simulation of the Avalanche Process in the G-APD and Circuitry Analysis of the SiPM does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with Simulation of the Avalanche Process in the G-APD and Circuitry Analysis of the SiPM, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Simulation of the Avalanche Process in the G-APD and Circuitry Analysis of the SiPM will most certainly appreciate the feedback.