Physics – Instrumentation and Detectors
Scientific paper
2007-09-07
Nucl.Instrum.Meth.A582:429-455,2007
Physics
Instrumentation and Detectors
Accepted by NIM A
Scientific paper
10.1016/j.nima.2007.08.157
The construction of the ATLAS electromagnetic liquid argon calorimeter modules is completed and all the modules are assembled and inserted in the cryostats. During the production period four barrel and three endcap modules were exposed to test beams in order to assess their performance, ascertain the production quality and reproducibility, and to scrutinize the complete energy reconstruction chain from the readout and calibration electronics to the signal and energy reconstruction. It was also possible to check the full Monte Carlo simulation of the calorimeter. The analysis of the uniformity, resolution and extraction of constant term is presented. Typical non-uniformities of 0.5% and typical global constant terms of 0.6% are measured for the barrel and end-cap modules.
Mandell L. i.
Stroynowsk R. i.
Aharrouche Mohamed
Aleksa Martin
Banfi D.
No associations
LandOfFree
Response Uniformity of the ATLAS Liquid Argon Electromagnetic Calorimeter does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Response Uniformity of the ATLAS Liquid Argon Electromagnetic Calorimeter, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Response Uniformity of the ATLAS Liquid Argon Electromagnetic Calorimeter will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-467808