Scientific paper
We report the results of laboratory tests of a pulse shape discrimination technique for reducing β--decay background in germanium detectors. The technique distinguishes single-site from multiple-site energy deposits in the detector through differences in the shape of the collected current pulse. At 816 keV, >~83% of single-site events can be rejected, while retaining ~70% of typical photon events.
Campbell Randy
Lin Robert P.
Luke Paul N.
Madden Norman W.
Pehl Richard H.
No associations
LandOfFree
Pulse shape discrimination for background reduction in germanium detectors does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Pulse shape discrimination for background reduction in germanium detectors, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Pulse shape discrimination for background reduction in germanium detectors will most certainly appreciate the feedback.