Physics
Scientific paper
Jul 2003
adsabs.harvard.edu/cgi-bin/nph-data_query?bibcode=2003icrc....2..821b&link_type=abstract
Proceedings of the 28th International Cosmic Ray Conference. July 31-August 7, 2003. Trukuba, Japan. Under the auspices of the I
Physics
Scientific paper
We have developed a fully automated test facility to evaluate 9 inch PMTs for Pierre-Auger Surface Detectors. The facility, now located at the Auger site in Malargue, Argentina, is capable of testing 16 PMTs simultaneously. It au¨ tomatically measures the following quantities: Gain (vs. High Voltage), Relative Quantum Efficiency, Single Photo electron Distribution, Excess Noise Factor, Dyno de/Ano de Ratio, Dark Pulse Rate, After Pulse Ratio, and Pulse Linearity. Using this facility, 441 PMTs have already been tested and evaluated, of which 95% have passed our specifications. These PMTs are being deployed in water tanks successfully. We plan to continuously test the remaining 4500 PMTs during the production deployment for the next three years.
Arisaka Katsushi
Barnhill David
García Beatriz
Jillings Chris
Ohnuki Tohru
No associations
LandOfFree
Production Test System and Results on Large PMTs for Pierre-Auger Surface Detectors does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Production Test System and Results on Large PMTs for Pierre-Auger Surface Detectors, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Production Test System and Results on Large PMTs for Pierre-Auger Surface Detectors will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-1136563