Physics – Instrumentation and Detectors
Scientific paper
2007-08-15
Nucl.Instrum.Meth.A582:569-574,2007
Physics
Instrumentation and Detectors
to be published in NIM A, corrected typos in Table 1 and Fig.6 of the previous version
Scientific paper
10.1016/j.nima.2007.08.180
We report the preparation of neutron-activated xenon for the calibration of liquid xenon (LXe) detectors. Gamma rays from the decay of xenon metastable states, produced by fast neutron activation, were detected and their activities measured in a LXe scintillation detector. Following a five-day activation of natural xenon gas with a Cf-252 (4 x 10^5 n/s) source, the activities of two gamma ray lines at 164 keV and 236 keV, from Xe-131m and Xe-129m metastable states, were measured at about 95 and 130 Bq/kg, respectively. We also observed three additional lines at 35 keV, 100 keV and 275 keV, which decay away within a few days. No long-lifetime activity was observed after the neutron activation.
Hasty R.
Kastens L.
Manzur A.
McKinsey Daniel N.
Ni Kaixuan
No associations
LandOfFree
Preparation of Neutron-activated Xenon for Liquid Xenon Detector Calibration does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Preparation of Neutron-activated Xenon for Liquid Xenon Detector Calibration, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Preparation of Neutron-activated Xenon for Liquid Xenon Detector Calibration will most certainly appreciate the feedback.