Physics – Instrumentation and Detectors
Scientific paper
2010-05-28
Nucl.Instrum.Meth.A622:225-230,2010
Physics
Instrumentation and Detectors
Scientific paper
10.1016/j.nima.2010.06.216
We present the timing characteristics of the flash ADC readout of the GlueX forward calorimeter, which depends on precise measurement of arrival time of pulses from FEU 84-3 photomultiplier tubes to suppress backgrounds. The tests presented were performed using two different 250 MHz prototype flash ADC devices, one with eight-bit and one with twelve-bit sampling depth. All measured time resolutions were better than 1 ns, independent of signal size, which is the design goal for the GlueX forward calorimeter. For pulses with an amplitude of 100 mV the timing resolution is 0.57 +- 0.18 ns, while for 500 mV pulses it is 0.24 +- 0.08 ns.
Bennett J. V.
Ito Makoto
Kornicer M.
Shepherd M. R.
No associations
LandOfFree
Precision timing measurement of phototube pulses using a flash analog-to-digital converter does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Precision timing measurement of phototube pulses using a flash analog-to-digital converter, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Precision timing measurement of phototube pulses using a flash analog-to-digital converter will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-16525