Crystal arrays made of LSO and LuAP crystals 2x2x10 mm pixels were manufactured for evaluation of detector with depth-of-interaction (DOI) determination capability intended for small animal positron emission tomograph. Position-sensitive LSO/LuAP phoswich DOI detector based on crystal 8x8 arrays and HAMAMATSU R5900-00-M64 position-sensitive multi-anode photomultiplier tube was developed and evaluated. Time resolution was found to be not worse than 1.0 ns FWHM for both layers, and spatial resolution mean value was 1.5 mm FWHM for the center of field-of-view.
Annenkov A.
Fedorov Arkady
Kholmetsky A.
Korzhik M.
Lecoq P.
No associations
LandOfFree
Position-Sensitive Detector with Depth-of-Interaction Determination for Small Animal PET does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Position-Sensitive Detector with Depth-of-Interaction Determination for Small Animal PET, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Position-Sensitive Detector with Depth-of-Interaction Determination for Small Animal PET will most certainly appreciate the feedback.