Multiple-GEM detectors are considered to be good candidates for tracking devices in experiments with high hadronic background. We present the results of the triple-GEM detectors beam test in a high intensity pion beam. The detectors had an optimized two-dimensional readout with minimized strip capacitance. Such optimization permitted the starting point of the efficiency plateau down to a gain of 5000. The probability of GEM discharges induced by heavily ionizing particles has been measured as a function of gain: at a gain of 20000 it amounts to 10^(-11) per incident particle. Such a value will ensure safe operation of the detector in the conditions of forward region of the LHC experiments.
Bondar Alex
Buzulutskov Alexei
Shekhtman L.
Sokolov Andrei
Vasiljev A.
No associations
LandOfFree
Performance of the triple-GEM detector with optimized 2-D readout in high intensity hadron beam does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Performance of the triple-GEM detector with optimized 2-D readout in high intensity hadron beam, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Performance of the triple-GEM detector with optimized 2-D readout in high intensity hadron beam will most certainly appreciate the feedback.