Physics – Nuclear Physics
Scientific paper
Feb 1998
adsabs.harvard.edu/cgi-bin/nph-data_query?bibcode=1998nuphs..61..289m&link_type=abstract
Nuclear Physics B, Proc. Suppl., Vol. 61B, p. 289 - 294
Physics
Nuclear Physics
Scientific paper
A transition radiation detector (TRD) consisting of three identical modules each of 12 m2 area, has been developed and installed on the MACRO detector at the Gran Sasso National Laboratory. The modular structure, based on single proportional tubes and polyethylene foam radiators, allows to cover a large area. The TRD measures the residual energies (up to 1 TeV) of penetrating muons at the MACRO depth. Calibrations from prototypes operated in a test beam at the CERN P.S. are quoted. The author presents preliminary data from the first TRD module, collected in a period of about one year.
No associations
LandOfFree
Performance of the transition radiation detector of the MACRO experiment. does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Performance of the transition radiation detector of the MACRO experiment., we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Performance of the transition radiation detector of the MACRO experiment. will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-1228081