The performance of prototypes for the ALICE electromagnetic sampling calorimeter has been studied in test beam measurements at FNAL and CERN. A $4\times4$ array of final design modules showed an energy resolution of about 11% /$\sqrt{E(\mathrm{GeV})}$ $\oplus$ 1.7 % with a uniformity of the response to electrons of 1% and a good linearity in the energy range from 10 to 100 GeV. The electromagnetic shower position resolution was found to be described by 1.5 mm $\oplus$ 5.3 mm /$\sqrt{E \mathrm{(GeV)}}$. For an electron identification efficiency of 90% a hadron rejection factor of $>600$ was obtained.
No associations
LandOfFree
Performance of prototypes for the ALICE electromagnetic calorimeter does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Performance of prototypes for the ALICE electromagnetic calorimeter, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Performance of prototypes for the ALICE electromagnetic calorimeter will most certainly appreciate the feedback.