The Tek 1024×1024 CCD system newly installed at Yunnam Observatory
is tested. According to the user's manual, the full well capacity of
this CCD should be 420.4K electrons/pixel (with the gain of 5), but the
real linear range (with non-linearity <1%) is less than 130K
electrons, in the authors preliminary experiments.
Lin Qing
Qin Dao
Yao Baoan
Zhang Chunsheng
No associations
LandOfFree
On the linearity of the Tek 1024 CCD at Yunnan Observatory. does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with On the linearity of the Tek 1024 CCD at Yunnan Observatory., we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and On the linearity of the Tek 1024 CCD at Yunnan Observatory. will most certainly appreciate the feedback.