Four position sensitive proportional counters will be delivered by the Danish Space Research Institute as focal plane instruments for the Soviet Danish Röntgen Telescope (SODART). These detectors will incorporate the novel microstrip electrode design. This design has been tested with xenon gas and an 55Fe X-ray source. The energy resolution is better than 14% FWHM up to gas gains of 104, the rise time of pulses from X-ray events is fast, 100 ns, allowing for efficient background rejection by rise time analysis. A position resolution of 1 mm FWHM is easily obtained by resistive charge division of the signals from the cathodes and from a wire grid placed in front of the microstrip plate. The detector performs without significant degradation of gain or energy resolution at fluxes exceeding 106/s cm2 and a total dose of 1011 photons.
Bahnsen Axel
Budtz-JØRgensen Carl
Jonasson Per
Madsen Morten
Oed Anton
No associations
LandOfFree
Microstrip proportional counters for X-ray astronomy does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Microstrip proportional counters for X-ray astronomy, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Microstrip proportional counters for X-ray astronomy will most certainly appreciate the feedback.