Physics – Instrumentation and Detectors
Scientific paper
2002-07-08
Physics
Instrumentation and Detectors
5 pages, 4 figures; Presented on a seminar at TRI/Princeton; To be published in Proceedings of the "Symposium: In Memory of Dr
Scientific paper
The main problem in liquid porosimetry, which prevents to see the pore sizes smaller than 2 microns in diameter, is direct gas diffusion flow through a micro-porous membrane. This diffusion causes bubbles formation below the membrane and that spoils extrusion (intrusion) data, as one cannot distinguish the volume of extrusion (intrusion) liquid from the volume of formatted bubbles. The suggested below method cures the problem by creating the liquid flow below the membrane. The flow washes out all of the small bubbles preventing them to grow. That allows using the membrane at higher differential pressures, even higher than minimum bubble point of the membrane, without spoiling data.
No associations
LandOfFree
Method for control gas diffusion and bubbles formation in liquid porosimetry does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Method for control gas diffusion and bubbles formation in liquid porosimetry, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Method for control gas diffusion and bubbles formation in liquid porosimetry will most certainly appreciate the feedback.