Superconducting tunnel junctions (STJ) have been developed mainly at high-resolution spectrometers for use in X-ray astronomy. A FWHM energy resolution of 112 eV at 5.9 keV is obtained using an STJ developed at Nippon Steel Corporation connected with a cooled FET (/~100 K). The pulse height of the signal is represented by a logarithmic function of energy based on consideration of the recombination of the quasi-particles in the junction. Experiments using series-connected STJs for an imaging radiation detector are performed. Both the pulse height and the rise time of signals from 241Am /α-particles indicate good position sensitivity with resolution less than 0.5 mm.

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

Measurements of the linearity of an STJ and position resolution of series-connected STJs does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with Measurements of the linearity of an STJ and position resolution of series-connected STJs, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Measurements of the linearity of an STJ and position resolution of series-connected STJs will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFWR-SCP-O-1830569

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.