Physics – Instrumentation and Detectors
Scientific paper
2009-06-16
Appl.Phys.Lett.94:183504,2009
Physics
Instrumentation and Detectors
3 pages; 3 figs
Scientific paper
10.1063/1.3127232
The thermal transport in partially trenched silicon nitride membranes has been studied in the temperature range from 0.3 to 0.6 K, with the transition edge sensor (TES), the sole source of membrane heating. The test configuration consisted of Mo/Au TESs lithographically defined on silicon nitride membranes 1 micron thick and 6 mm^2 in size. Trenches with variable depth were incorporated between the TES and the silicon frame in order to manage the thermal transport. It was shown that sharp features in the membrane surface, such as trenches, significantly impede the modes of phonon transport. A nonlinear dependence of thermal resistance on trench depth was observed. Partial perforation of silicon nitride membranes to control thermal transport could be useful in fabricating mechanically robust detector devices.
Bleem Lindsey
Carlstrom John E.
Chang Chulhoon
Crites Abigail
Datesman Aaron
No associations
LandOfFree
Low Temperature Thermal Transport in Partially Perforated Silicon Nitride Membranes does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Low Temperature Thermal Transport in Partially Perforated Silicon Nitride Membranes, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Low Temperature Thermal Transport in Partially Perforated Silicon Nitride Membranes will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-475798