Leakage Tests for Am-241 Solid Sources Used for Liquid Xenon Detector Monitoring

Physics – Medical Physics

Scientific paper

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

5 pages, 1 figure, 3 tables

Scientific paper

10.1016/j.nima.2006.02.170

Experimental tests are described to check solid Am-241 radioactive sources used to monitor operation of liquid xenon detectors. In particular, radioactive leakage was tested in extreme temperature conditions following immersion in liquid nitrogen for different time periods. No radioactivity loss was detected from the sources considered. The paper describes the source characteristics, test methods and results obtained.

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

Leakage Tests for Am-241 Solid Sources Used for Liquid Xenon Detector Monitoring does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with Leakage Tests for Am-241 Solid Sources Used for Liquid Xenon Detector Monitoring, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Leakage Tests for Am-241 Solid Sources Used for Liquid Xenon Detector Monitoring will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFWR-SCP-O-704002

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.