Experimental tests are described to check solid Am-241 radioactive sources used to monitor operation of liquid xenon detectors. In particular, radioactive leakage was tested in extreme temperature conditions following immersion in liquid nitrogen for different time periods. No radioactivity loss was detected from the sources considered. The paper describes the source characteristics, test methods and results obtained.
Bazzarri S.
Cicoli G.
Felice P. de
Rossi Giancarlo
Sedda F.
No associations
LandOfFree
Leakage Tests for Am-241 Solid Sources Used for Liquid Xenon Detector Monitoring does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Leakage Tests for Am-241 Solid Sources Used for Liquid Xenon Detector Monitoring, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Leakage Tests for Am-241 Solid Sources Used for Liquid Xenon Detector Monitoring will most certainly appreciate the feedback.