Large-area CdTe diode detector for space application

Physics

Scientific paper

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

8

Scientific paper

The current status of Schottky CdTe diode detectors, especially in view of their space application for hard X-ray and gamma-ray astronomy, are reported. For practical use in space science, a large-area CdTe diode with a size of 21.5×21.5mm2 and a thickness of 0.5mm was developed. A good energy resolution, 2.8keV (FWHM) at -20°C, and high homogeneity to within 0.2% over the detector were achieved for the spectral performance. This device has successfully passed a series of tests required for its use in space, in view of utilizing Japanese M-V rockets. The tests include the mechanical environment test, vacuum test, long run for weeks and proton-beam radiation. Initial results from a 2×2 segmented electrode large-area device with a guard-ring are also presented.

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

Large-area CdTe diode detector for space application does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with Large-area CdTe diode detector for space application, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Large-area CdTe diode detector for space application will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFWR-SCP-O-1034634

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.