Scientific paper
Design, construction, and sensitivity estimates for the performance of proportional counters customarily neglect the consequences of the finite path length of electrons resulting from the primary interaction between the incident x-ray and the proportional counter gas. In many situations this can lead to errors in response functions as a result of ignoring 'charge sharing' between adjacent proportional counter cells which are held in anti-coincidence, or charge lost by the electrons escaping the gas volume into the detector window or into inactive regions. We illustrate the importance of these effects with Monte-Carlo simulations.
Austin Robert A.
Dietz Kurtis L.
O'Dell Stephen L.
Ramsey Brian D.
Tennant Alan
No associations
LandOfFree
Importance of electron tracking in understanding the response of proportional counters does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Importance of electron tracking in understanding the response of proportional counters, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Importance of electron tracking in understanding the response of proportional counters will most certainly appreciate the feedback.