Purpose : To relate the physical transmittance parameters of the water equivalent Gafchromic EBT 2 Film with the delivered dose in a transparent absolute calibration protocol. The protocol should be easy to understand, easy to perform, and should be able to predict the residual dose error. Conclussions : The gafchromic EBT2 Films are properly calibrated with an accessible robust calibration protocol. The protocol largely deals with the uniformity problems of the Film. The proposed method allowed to relate the dose with the red channel transmittance using only T0, T_inf, and a dose scaling factor. Based on the local and global uniformity the red channel dose errors could be predicted to be smaller than 5%.
Crijns Wouter
den Heuvel Frank Van
Maes Frederik
No associations
LandOfFree
Gafchromic film dosimetry: calibration methodology and error analysis does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Gafchromic film dosimetry: calibration methodology and error analysis, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Gafchromic film dosimetry: calibration methodology and error analysis will most certainly appreciate the feedback.