First Measures of Sensitivity for the Post-COSTAR GHRS, With Interim Values for Data Analysis

Physics

Scientific paper

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

Hubble Space Telescope, Hst, Space Telescope Science Institute, Ghrs, Goddard High Resolution Spectograph

Scientific paper

Observations were made with the GHRS through both apertures after the COSTAR mirrors were deployed to evaluate changes in instrument sensitivity. The ratio of post- to pre-COSTAR count rates are examined and illustrated for a series of first-order grating observations and an OSCAN with Echelle-B. These ratios behave about as expected for wavelength > 2500A (about 1 for the LSA; about 2 for the SSA) but show a broad deficit in response from about 1500 to 2500A, as well as a peak in response near 1300A. These features are very similar to what has been seen with the blue side of the Faint Object Spectrograph, although the underlying origins are not yet understood.
A full ab initio determination of GHRS Sensitivity will be made early in Cycle 4. As an interim measure for science data calibration we provide here a relationship between pre-COSTAR sensitivities and those for the post-COSTAR instrument. The ratios of post- to pre- COSTAR count rates for both the LSA and SSA have been fitted with functions. These functions have been applied to the pre-COSTAR sensitivities that are used to flux calibrate science observations, and copies have been posted on STEIS and delivered to the Calibration Data Base System (CDBS).

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

First Measures of Sensitivity for the Post-COSTAR GHRS, With Interim Values for Data Analysis does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with First Measures of Sensitivity for the Post-COSTAR GHRS, With Interim Values for Data Analysis, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and First Measures of Sensitivity for the Post-COSTAR GHRS, With Interim Values for Data Analysis will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFWR-SCP-O-1024247

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.