Physics
Scientific paper
Jun 1998
adsabs.harvard.edu/cgi-bin/nph-data_query?bibcode=1998aipc..430..428r&link_type=abstract
The eleventh international conference on fourier transform spectroscopy. AIP Conference Proceedings, Volume 430, pp. 428-431 (1
Physics
1
Infrared Spectrometers, Auxiliary Equipment, And Techniques, Photodetectors
Scientific paper
For strongly absorbing bands measured with a Fourier transform infrared (FT-IR) spectrometer, the effects of a nonlinear detector response must be eliminated before Beer's law linearity can be achieved. The effect of the non-linear response of mercury cadmium telluride (MCT) detectors has been evaluated on three commercial FT-IR spectrometers. The greater the photon flux, photon flux density, and the smaller the area of the detector on which the source image is focused, the greater are the effects of the nonlinearity. Detector nonlinearity is manifested by the generation of nonlinear Beer's law plots. A simple correction algorithm has been applied to Beer's law data acquired using both photoconductive MCT and pyroelectric detectors and found to work well.
Griffiths Peter R.
Richardson Robert L.
Yang Husheng
No associations
LandOfFree
Factors inducing and correction of photometric error introduced to FT-IR spectrometers by a nonlinear detector response does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Factors inducing and correction of photometric error introduced to FT-IR spectrometers by a nonlinear detector response, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Factors inducing and correction of photometric error introduced to FT-IR spectrometers by a nonlinear detector response will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-1330506