A p-type HPGe detector has been segmented by cutting, with a diamond saw, and etching six circumferential grooves through the Li-diffused dead layer. The degree of segmentation was tested with the 88 keV gamma rays from a well-collimated source of 109Cd. The fraction of events, recognized as occurring in more than one segment, and rejected from the energy interval 2038 +/- 5 keV, was measured as 0.59.
Avignone III Frank T.
Cox Christopher E.
Hossbach Todd W.
Jennings Wayne
King III George S.
No associations
LandOfFree
Fabrication and Performance Tests of a Segmented P-Type HPGe Detector does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Fabrication and Performance Tests of a Segmented P-Type HPGe Detector, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Fabrication and Performance Tests of a Segmented P-Type HPGe Detector will most certainly appreciate the feedback.