Physics – Instrumentation and Detectors
Scientific paper
2007-09-21
JINST2:P12003,2007
Physics
Instrumentation and Detectors
12 pages, 10 figures, 1 table, submitted to Journal of Instrumentation
Scientific paper
10.1088/1748-0221/2/12/P12003
We present methods to improve the energy resolution of single-channel, room-temperature Cadmium-Zinc-Telluride (CZT) detectors. A new preamplifier design enables the acquisition of the actual transient current from the crystals and straightforward data analysis methods yield unprecedented energy resolution for our test-detectors. These consist of an eV-CAPture Plus crystal as standard and 1 cm cube Frisch collar crystals created in-house from low-grade coplanar grid detectors. Energy resolutions of 1.9% for our collar detectors and 0.8% for the eV crystal at 662 keV were obtained. The latter compares favourably to the best existing energy resolution results from pixel detectors.
Ramachers Y.
Stewart D. Y.
No associations
LandOfFree
Energy resolution improvement in room-temperature CZT detectors does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Energy resolution improvement in room-temperature CZT detectors, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Energy resolution improvement in room-temperature CZT detectors will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-257084