Energy resolution and energy-light response of CsI(TI) scintillators for charged particle detection

Physics – Nuclear Physics – Nuclear Experiment

Scientific paper

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

24 pages, 7 figures

Scientific paper

10.1016/S0168-9002(00)00542-8

This article describes the crystal selection and quality control utilized to develop and calibrate a high resolution array of CsI(TI) scintillator crystals for the detection of energetic charged particles. Alpha sources are used to test the light output variation due to thallium doping gradients. Selection of crystals with better than 1% non-uniformity in light output is accomplished using this method. Tests with 240 MeV alpha beam reveal that local light output variations within each of the tested CsI(TI) crystals limit the resolution to about 0.5%. Charge and mass dependences in the energy - light output relationship are determined by calibrating with energetic projectile fragmentation beams.

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

Energy resolution and energy-light response of CsI(TI) scintillators for charged particle detection does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with Energy resolution and energy-light response of CsI(TI) scintillators for charged particle detection, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Energy resolution and energy-light response of CsI(TI) scintillators for charged particle detection will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFWR-SCP-O-398325

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.