Physics – Instrumentation and Detectors
Scientific paper
1997-10-08
Nucl.Instrum.Meth. A403 (1998) 425
Physics
Instrumentation and Detectors
10 pages, 4 PostScript figures, uses elsart.sty, submitted to Nucl. Instrum. Meth. A
Scientific paper
10.1016/S0168-9002(97)01152-2
Cathode electrodes of the Si(Li) detector elements of the SIXA X-ray spectrometer array are formed by gold-palladium alloy contact layers. The equivalent thickness of gold in one element was measured by observing the characteristic L-shell X-rays of gold excited by monochromatised synchrotron radiation with photon energies above the L3 absorption edge of gold. The results obtained at 4 different photon energies below the L2 edge yield an average value of 22.4(35) nm which is consistent with the earlier result extracted from detection efficiency measurements. PACS: 29.40.Wk; 85.30.De; 07.85.Nc; 95.55.Ka Keywords: Si(Li) detectors, X-ray spectrometers, X-ray fluorescence, detector calibration, gold electrodes, synchrotron radiation
Hämäläinen Keijo
Huotari Simo
Tikkanen Tuomo
No associations
LandOfFree
Electrode thickness measurement of a Si(Li) detector for the SIXA array does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Electrode thickness measurement of a Si(Li) detector for the SIXA array, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Electrode thickness measurement of a Si(Li) detector for the SIXA array will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-130121