Significant systematic errors in high-precision Penning trap mass spectrometry can result from electric and magnetic field imperfections. An experimental procedure to minimize these uncertainties is presented for the on-line Penning trap mass spectrometer ISOLTRAP, located at ISOLDE/CERN. The deviations from the ideal magnetic and electric fields are probed by measuring the cyclotron frequency and the reduced cyclotron frequency, respectively, of stored ions as a function of the time between the ejection of ions from the preparation trap and their capture in the precision trap, which influences the energy of their axial motion. The correction parameters are adjusted to minimize the frequency shifts.
Beck David
Blaum Klaus
Bollen G.
Delahaye P.
George Samuel
No associations
LandOfFree
Electric and magnetic field optimization procedure for Penning trap mass spectrometers does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Electric and magnetic field optimization procedure for Penning trap mass spectrometers, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Electric and magnetic field optimization procedure for Penning trap mass spectrometers will most certainly appreciate the feedback.