We evaluate the frequency error from distributed cavity phase in the caesium fountain clock PTB-CSF2 at the Physikalisch-Technische Bundesanstalt with a combination of frequency measurements and ab initio calculations. The associated uncertainty is 1.3E-16, with a frequency bias of 0.4E-16. The agreement between the measurements and calculations explains the previously observed frequency shifts at elevated microwave amplitude. We also evaluate the frequency bias and uncertainty due to the microwave lensing of the atomic wavepackets. We report a total PTB-CSF2 systematic uncertainty of 4.1E-16.
Gerginov Vladislav
Gibble Kurt
Li Renwei
Nemitz Nils
Weyers Stefan
No associations
LandOfFree
Distributed cavity phase frequency shifts of the caesium fountain PTB-CSF2 does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Distributed cavity phase frequency shifts of the caesium fountain PTB-CSF2, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Distributed cavity phase frequency shifts of the caesium fountain PTB-CSF2 will most certainly appreciate the feedback.