An algorithm to measure the drift time of charge carriers in p-type point contact (PPC) high-purity germanium (HPGe) detectors from the signals processed with a charge-sensitive preamplifier is introduced. It is demonstrated that the drift times can be used to estimate the distance of charge depositions from the point contact and to characterize losses due to charge trapping. A correction for charge trapping effects over a wide range of energies is implemented using the measured drift times and is shown to improve the energy resolution by up to 30%.
Amman Mark
Chan Yuen-Dat
Detwiler Jason A.
Loach J. C.
Looker Q.
No associations
LandOfFree
Determining the Drift Time of Charge Carriers in P-Type Point-Contact HPGe Detectors does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Determining the Drift Time of Charge Carriers in P-Type Point-Contact HPGe Detectors, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Determining the Drift Time of Charge Carriers in P-Type Point-Contact HPGe Detectors will most certainly appreciate the feedback.